Code on display E A&D HR-A

Overload

The load on the weighing pan exceeds the balance's weighing capacity, so no value is shown. A load above capacity strains the precision weighing mechanism.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Remove the sample from the weighing pan immediately.
  2. Weigh the sample in portions so that it stays within capacity.
  3. If necessary, use a lighter container.
  • Do not press on the weighing pan or leave a load above capacity on it.

Possible causes

  • Placing a sample heavier than capacity
  • Using a heavy container with the sample

Do you need service?

The display returns to normal once the load is removed. If E also appears with an empty weighing pan, contact service for a load cell check.

Models showing this code: HR-A