Weighing pan error
The measured value is much lower than it should be. The balance detects that the weighing pan is missing or installed incorrectly. Weighings made without the pan seated correctly are not reliable.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Check that the weighing pan (and the pan support, if any) is installed correctly and does not touch anything.
- Remove the weighing pan and reinstall it correctly.
- If the error persists, calibrate the balance (perform a span adjustment).
Possible causes
- Weighing pan or pan support not installed
- Weighing pan seated incorrectly or touching the draft shield
- Calibration drift
Do you need service?
If -E keeps appearing even though the weighing pan is installed correctly and calibration has been done, the weighing sensor must be checked.
Models showing this code: GX
Source: A&D official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.