Code on display Error 7 A&D GX

Internal weight mechanism error

Shown on the display as Error 7: the mechanism that loads and removes the internal calibration weight is not working properly. Calibration with the internal weight cannot be performed in this case.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Empty the weighing pan and restart the procedure from the beginning.
  2. Turn the balance off and on again, then retry internal calibration.
  • Do not open the device to interfere with the mechanism.

Possible causes

  • Internal weight mechanism jamming
  • Mechanism problem after transport
  • Shaking during calibration

Do you need service?

If Error 7 persists when the procedure is repeated, the internal weight mechanism is faulty and must be repaired by technical service.

Models showing this code: GX

Source: A&D official user manual · Open manufacturer document