Internal weight error
The balance did not get the expected value while measuring with the internal calibration weight. The manual advises checking that the pan is empty.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Check that the pan is completely empty.
- Close the draft shield doors.
- Repeat the calibration or calibration test.
Possible causes
- Load on the pan during calibration
- Vibration or drafts
- Internal weight mechanism fault
Do you need service?
If the error continues with an empty pan, the internal weight mechanism must be checked by a technician.
Source: A&D official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.