Code on display EC,E00 A&D GR

Communication protocol error

The balance could not decode the data received over RS-232C. This code is sent to the connected device, not shown on the balance display.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Set the same baud rate on the balance and the PC.
  2. Compare the parity and data bit settings.
  3. Unplug and reconnect the RS-232C cable.

Possible causes

  • Baud rate mismatch
  • Different parity or data format
  • Loose cable connection

Do you need service?

If the error continues with matching settings, the interface board must be checked by a technician.

Source: A&D official user manual · Open manufacturer document