Code on display CH 0 A&D GR

Pan not empty, ARA not done

ARA adjusts the balance response speed to the environment. The pan must be empty for this.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Remove everything from the pan.
  2. Press the CAL key to return to weighing mode.
  3. Start ARA again.

Possible causes

  • Sample or container left on the pan
  • Object touching the pan

Do you need service?

This is solved by the user, no service is needed.

Source: A&D official user manual · Open manufacturer document