Code on display CAL E A&D EK-L

Calibration weight too heavy

The scale cannot match the weight on the pan to the expected calibration value. The calibration is not saved.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Check the weighing pan and the calibration weight.
  2. Press the CAL switch to return to weighing mode or switch the scale off with ON/OFF.
  3. Calibrate again with the correct weight.

Possible causes

  • Wrong nominal weight used
  • Extra object on the pan

Do you need service?

If CAL E appears with the correct weight, the balance adjustment must be checked by a technician.

Models showing this code: EK-15KL, EK-30KL

Source: A&D official user manual · Open manufacturer document