Code on display Error 5 A&D EK-L

Excessive load detected

The scale sees a force on the pan side far above normal. The manual asks you to confirm that the weighing pan is installed correctly.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Remove all items from the weighing pan.
  2. Check that the weighing pan is installed correctly.
  3. Switch the balance off and on again.
  • Do not place a load above capacity on the weighing pan.

Possible causes

  • Heavy objects on the pan
  • Pan installed incorrectly
  • External pressure on the pan

Do you need service?

If Error 5 remains with an empty, correctly installed pan, the load cell must be checked by a technician.

Models showing this code: EK-15KL, EK-30KL

Source: A&D official user manual · Open manufacturer document