Code on display Err 47 Precisa 165 BJ

Too many values for statistics

The number of values that can be stored in the statistics application was exceeded. A new value cannot be added to the statistics.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Print or note down the current statistics results.
  2. Clear the statistics memory and start a new series.

Possible causes

  • Statistics memory full

Do you need service?

New values can be added once the memory is cleared.

Models showing this code: 165 BJ

Source: Precisa official user manual · Open manufacturer document