Code on display UL Precisa Series 160 BJ

Below weighing range / pan missing

The measured value is below the device's weighing range. The manual gives a missing pan or pan holder as the cause.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Install the pan holder and the pan.
  2. Check that the pan is flat and seated correctly.
  3. Switch the balance off and on again.

Possible causes

  • The weighing pan is not installed
  • Pan holder not installed
  • The weighing pan is seated incorrectly

Do you need service?

If UL persists with the pan and its holder installed correctly, the measuring system must be checked by technical service.

Source: Precisa official user manual · Open manufacturer document