Process timeout
The scale waits for a steady value for the operation. If the value does not settle because of drafts or vibration, the operation is not completed.
You can usually fix this yourself.
What should you do?
Try the steps in order.
- Move the balance to a place away from drafts and vibration.
- Close the draft shield doors.
- Level the balance using the level indicator and repeat the operation.
Possible causes
- Air drafts
- Vibration
- Uneven table
Do you need service?
If timeouts keep occurring even in a suitable environment, a service check is required.
Source: OHAUS official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.