Code on display ERR 340 Minebea Intec Combics 3

Operating parameter (EEPROM) error

There is an inconsistency in the memory that holds the setting data. The manual recommends first turning the device off and on, and contacting service if the error continues.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Turn the device off.
  2. Turn it back on after a few seconds and check whether the code has disappeared.

Possible causes

  • EEPROM data corruption
  • Power outage

Do you need service?

If ERR 340 still appears after restarting, contact technical service to have the parameters loaded.

Models showing this code: CAISL3, CAIS3, CAISL3-U, CAIS3-U

Source: Minebea Intec official user manual