Code on display -OL- Mecmesin BFG

Momentary overload

A force greater than 120% of full-scale capacity is currently being applied to the load cell. If this continues, the sensor may be permanently damaged.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Remove the applied force immediately.
  2. Check that the indicator returns to zero.
  3. Place the device without accessories on a flat surface, turn it on while holding the MAX key, and read the load cell offset percentage. If it is above 5%, contact your supplier.
  • Do not trust the measurements of a device that has been overloaded. Consult your supplier.

Possible causes

  • Force applied above capacity
  • Force applied as a sudden impact

Do you need service?

If the indicator does not return to zero after the force is removed, the load cell may be damaged. technical service should inspect it.

Models showing this code: BFG

Source: Mecmesin official user manual