Code on display ERR H KERN & Sohn ABS

Hardware error

The manual describes this message as a general hardware error. If it is still shown after a restart, the fault is permanent and the balance should not be used.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Switch the balance off with the ON/OFF key.
  2. Unplug the power adapter and wait a few seconds.
  3. Plug the adapter back in and switch the balance on again.

Possible causes

  • Fault on the electronic board
  • Brief power interruption or fluctuation

Do you need service?

If ERR H remains after a restart, the electronic board must be checked by a technician.

Source: KERN & Sohn official user manual · Open manufacturer document