Code on display Unst Dikomsan KS-06

Initial stability error

The scale could not read a stable value while setting the zero point at power-on. Instability at startup affects the accuracy of later weighings.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Make sure the pan is empty and correctly seated.
  2. Move the balance to a level, solid, vibration-free surface and away from air drafts.
  3. Switch the balance off and on again.

Possible causes

  • Air drafts
  • Vibrating or sloping floor
  • There was a load on the weighing pan at switch-on

Do you need service?

If startup instability persists in a suitable environment, the load cell or electronics board must be checked by service.

Models showing this code: KS-06

Source: Dikomsan official user manual