Code on display --OL-- Citizen CX Series

Overload

The weight on the weighing pan is above the weighing range, so no value is displayed. On precision balances, continuous overload can damage the measuring system.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Remove the excess weight from the weighing pan.
  2. Weigh the sample in portions that fit the capacity.
  • Do not place shock loads or loads above capacity on the weighing pan.

Possible causes

  • Sample heavier than capacity
  • An extra container or object was placed on the weighing pan

Do you need service?

If --OL-- also appears with an empty pan, contact technical service to check the measuring cell.

Models showing this code: CX 65, CX 120, CX 220, CX 301

Source: Citizen official user manual