Code on display Error 9 Citizen CX Series

Real-time clock error

Because the clock circuit that keeps the date and time is not working, correct timestamps cannot be produced on GLP printouts. The manual says to contact the service center.

This fault needs a technician.

What should you do?

You can safely do these while waiting for service.

  1. Turn the balance off and on again. If the error persists, contact service.
  • Do not open the device to replace the backup battery.

Possible causes

  • RTC circuit fault
  • Clock backup battery depleted

Do you need service?

Because the clock circuit is on the board, its repair must be done by technical service.

Models showing this code: CX 65, CX 120, CX 220, CX 301

Source: Citizen official user manual