Code on display Error 2 Citizen CG Series

Pan loaded at startup

According to the manual, this occurs at balance startup. Because there is a load on the pan during startup, the zero point and calibration cannot be started correctly.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Turn the balance off.
  2. Empty the weighing pan completely.
  3. Turn the balance on again with the pan empty.

Possible causes

  • Sample or container on the pan at power-on
  • Object forgotten on the pan

Do you need service?

It is resolved by restarting with an empty pan. No service is required.

Source: Citizen official user manual