Code on display Err 3 Baykon BX2

Calibration cannot be performed

The device cannot be calibrated because no valid signal is received from the load cell during calibration. According to the manual, the load cell or its cable may be faulty.

This fault needs a technician.

What should you do?

You can safely do these while waiting for service.

  1. Visually check the load cell cable for crushing, cuts or a disconnected connector.
  2. Turn the device off, wait 30 seconds and turn it on again. If the error continues, contact service.
  • Do not open the indicator housing or tamper with the board and connections.
  • Do not touch the calibration jumper and the legal seal.

Possible causes

  • Load cell fault
  • Load cell cable broken or crushed
  • Loose connection terminals

Do you need service?

Measuring the load cell and its cable and then calibrating must be done by technical service.

Models showing this code: BX2

Source: Baykon official user manual · Open manufacturer document