Overload
The H message means the load on the weighing pan exceeds the scale's maximum capacity. Overload can damage the load cell, so remove the load immediately. If H also appears with an empty pan, there may be mechanical damage.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Remove the load from the weighing pan immediately.
- Check that the load to be weighed is below the scale's maximum capacity.
- With the pan empty, turn the balance off and on again.
- Do not place loads that exceed capacity or shock loads on the weighing pan.
Possible causes
- Load exceeding capacity
- Shock loading on the weighing pan
- Mechanical damage to the load cell
Do you need service?
If the H message also appears with an empty pan, the load cell may be damaged. Contact technical service.
Models showing this code: ME-01/A, ME-01/N, ME-01/A/LCD, ME-01/N/LCD
Source: AXIS official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.