Code on display ------ AXIS ATN

Taring could not be completed

Dashes (------) on the display mean the scale could not finish the tare/zeroing process. The scale completes it only when it gets a stable reading. Vibration, air currents, or damage can prevent this.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Place the scale on a sturdy, vibration-free table.
  2. Prevent air currents (doors, windows, air conditioning) and close the draft shield doors.
  3. Wait a few seconds and press the tare key again.

Possible causes

  • Unstable surface or vibration
  • Air drafts
  • Damage to the scale

Do you need service?

If taring still does not complete in a stable environment, the manual says to contact service.

Source: AXIS official user manual · Open manufacturer document