Code on display Err-H AXIS AGN/C

Load on the weighing pan at startup

The scale detected a load on the weighing pan while turning on and could not take the initial zero point. It usually clears when the weighing pan is emptied and the device is turned on again.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Empty the weighing pan completely.
  2. Switch the device off and on again.

Possible causes

  • Load or container on the weighing pan at startup
  • Mechanical damage to the scale

Do you need service?

If Err-H also appears with an empty weighing pan, contact technical service for a mechanical or electronic check of the scale.

Models showing this code: AGN50C, AGN100C, AGN200C

Source: AXIS official user manual