Code on display CAL E A&D MS70

Calibration weight too heavy

According to the manual, something touching the pan, the weight touching the heater cover glass or a wrong weight value causes this. The unit returns to weighing mode after 15 seconds.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Check the value of the calibration weight.
  2. Make sure the weight does not touch the heater cover glass.
  3. Check that nothing touches the pan.
  4. Repeat the calibration.

Possible causes

  • Calibration weight of the wrong value
  • Weight touching the cover glass
  • Something else touching the pan

Do you need service?

If the error continues with the correct weight, the weighing sensor must be checked by a technician.

Source: A&D official user manual · Open manufacturer document