Calibration weight too heavy
According to the manual, something touching the pan, the weight touching the heater cover glass or a wrong weight value causes this. The unit returns to weighing mode after 15 seconds.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Check the value of the calibration weight.
- Make sure the weight does not touch the heater cover glass.
- Check that nothing touches the pan.
- Repeat the calibration.
Possible causes
- Calibration weight of the wrong value
- Weight touching the cover glass
- Something else touching the pan
Do you need service?
If the error continues with the correct weight, the weighing sensor must be checked by a technician.
Source: A&D official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.