Code on display tilt A&D GX-L

Level check (balance not level)

The balance warns that it is not horizontal according to the spirit level on its body. A tilted balance cannot weigh accurately.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Look at the spirit level on the body.
  2. Turn the adjustable feet to bring the bubble inside the center circle.
  3. Press any key other than ON:OFF to dismiss the warning.

Possible causes

  • Balance standing on an uneven surface
  • Adjustable feet unbalanced
  • Balance has been moved

Do you need service?

This is not a fault. Leveling the balance with its feet is enough.

Models showing this code: GX-L

Source: A&D official user manual · Open manufacturer document