ECL repeatability error
In the self check performed with the electronically controlled load (ECL), the standard deviation of repeatability exceeded 50 divisions (50 d). This indicates that the installation environment is disturbing measurement stability.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Place the balance on a solid, vibration-free table away from drafts.
- Allow enough time for the balance to warm up.
- Run the self check again.
Possible causes
- Vibration or air drafts
- Balance standing on an unstable surface
- Temperature change
Do you need service?
If SD Error persists in the self check even after the environment is corrected, the balance's weighing performance should be checked by service.
Models showing this code: GX-L
Source: A&D official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.