Calibration weight too heavy
Shown on the display as CAL E: the weight used in the span adjustment (external calibration) is too heavy compared with the value the device expects. If calibration is performed with the wrong weight, all weighings will be inaccurate, so the procedure is not accepted.
You can usually fix this yourself.
What should you do?
Try the steps in order.
- Press the CAL key to return to weighing mode.
- Check that the weighing pan is installed correctly and that nothing else is on it.
- Check the value of the calibration weight you are using and the calibration weight value set on the balance.
- Repeat the calibration with the correct weight.
Possible causes
- Using a calibration weight heavier than the set value
- Calibration weight value set incorrectly
- Weighing pan installed incorrectly or another load on the pan
Do you need service?
If CAL E persists even with calibration using the correct, certified weight, the balance may need to be adjusted by service.
Models showing this code: GX-L
Source: A&D official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.