Stability error
This error, also shown on the display as Error 1, indicates that the weighing value could not be stabilized, so operations such as zeroing and calibration cannot be performed. The cause is usually environmental: vibration, air currents, temperature changes, static electricity, or a magnetic field.
You can usually fix this yourself.
What should you do?
Try the steps in order.
- Check around the weighing pan and make sure it does not touch the draft shield or any other object.
- Move the balance to a sturdy, vibration-free table away from air currents.
- Take the necessary precautions for statically charged or magnetic samples (e.g. an anti-static device, a suitable container).
- Press the CAL key to return to weighing mode and repeat the procedure.
Possible causes
- Vibration or shaking
- Air currents (air conditioner, door, window)
- Static electricity or a magnetic sample
- The weighing pan touching something
Do you need service?
According to the manual, the cause is the environmental conditions. If the error keeps recurring after the environment is corrected, a service check of the balance may be needed.
Models showing this code: GF
Source: A&D official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.