Code on display Lo A&D FZ-iWP

Sample weight too low

The balance cannot store the sample you want to save as a reference in counting (PCS) or percent (%) mode because it is too light. A reference that is too light makes counting and percent results unreliable.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Use a heavier reference sample (place more pieces in counting mode).
  2. With the weighing pan empty, press the RE-ZERO key to zero.
  3. Repeat the reference storage.

Possible causes

  • Reference sample too light
  • Too few pieces used for the reference

Do you need service?

The warning clears when storage is repeated with a heavier reference sample. Service is not required.

Models showing this code: FZ-iWP

Source: A&D official user manual · Open manufacturer document