Weighing stability error
The balance cannot stabilize the reading because of an environmental factor and cannot complete the process (for example calibration or weighing by command). Vibration, air drafts, temperature changes, static electricity or magnetic fields cause the value to keep fluctuating.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Press the CAL key to return to weighing mode.
- Move the balance to a sturdy table free of vibration and away from air drafts.
- Close the draft shield doors if present, and keep statically charged or magnetic samples away from the weighing pan.
- Set the Response Adjustment to suit the environment and repeat the process.
Possible causes
- Vibration on the table or floor
- Air drafts from an air conditioner, door or window
- Sudden temperature change
- Static electricity or magnetic field in the sample
Do you need service?
If the balance still cannot give a stable reading after the environmental conditions are corrected, contact technical service to check the load cell and electronics.
Models showing this code: FZ-iWP
Source: A&D official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.