Code on display EC,E06 A&D FZ-iWP

Command format error

The command sent contains invalid data, for example a numeric value that is not written in a valid format. The balance does not process the command.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Compare the data part of the command with the examples in the manual.
  2. Write the value in the correct format and send the command again.

Possible causes

  • Numeric value in the command written incorrectly
  • Decimal separator or digit error

Do you need service?

The error clears when the command data is sent in the correct format. There is no fault in the balance.

Models showing this code: FZ-iWP

Source: A&D official user manual · Open manufacturer document