Code on display EC,E03 A&D FZ-iWP

Communication timeout

With the timeout setting at t-UP1, the balance did not receive the next character of a command within one second. The command is not processed because it is incomplete.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Check the connection cable and connectors.
  2. Verify that the whole command is sent at once, together with the line terminator (CR LF).
  3. Retry communication.

Possible causes

  • Command characters sent slowly or in pieces
  • Connection cable problem or interrupted communication

Do you need service?

The error clears when the command is sent completely and on time. There is no fault in the balance.

Models showing this code: FZ-iWP

Source: A&D official user manual · Open manufacturer document