Code on display EC,E02 A&D FZ-iWP

Device not ready to process command

The balance cannot process the command it received at that moment. For example, it may have received a Q command while not in weighing mode or while a RE-ZERO operation was in progress.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Make sure the balance is in weighing mode.
  2. Increase the wait (delay) time between commands in the software.
  3. Send the command again.

Possible causes

  • Command sent while the balance is not in weighing mode
  • New command sent before the previous command (e.g. RE-ZERO) finished

Do you need service?

The error clears once the command timing is adjusted. No service is needed.

Models showing this code: FZ-iWP

Source: A&D official user manual · Open manufacturer document