Code on display -CAL E A&D FZ-i

Calibration weight too light

In calibration with an external weight, the weight placed on the weighing pan measured too light compared with the expected calibration value. The balance does not accept the calibration to prevent an incorrect adjustment.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Press the CAL key to return to weighing mode.
  2. Check that the weighing pan and pan support are installed correctly.
  3. Verify that the weight you are using matches the calibration value on the display.
  4. Repeat the calibration with the correct weight.

Possible causes

  • Using a weight lighter than the calibration value on the display
  • Weighing pan or pan support not installed or seated incorrectly
  • Weight not fully placed on the weighing pan

Do you need service?

If -CAL E continues with the correct weight and a correctly installed weighing pan, technical service must check the balance's adjustment and load cell.

Models showing this code: FZ-i

Source: A&D official user manual · Open manufacturer document