Code on display EC,E20 A&D FX-GD

Calibration weight too heavy

Shown on the display as CAL E: the weight used for external calibration is too heavy compared with the value the device expects. If calibration is performed with the wrong weight, all weighings will be inaccurate, so the procedure is not accepted.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Press the CAL key to return to weighing mode.
  2. Check that the weighing pan is installed correctly and that nothing else is on it.
  3. Check the value of the calibration weight you are using and the calibration weight value set on the balance.
  4. Repeat the calibration with the correct weight.

Possible causes

  • Using a calibration weight heavier than the set value
  • Calibration weight value set incorrectly
  • Weighing pan installed incorrectly or another load on the pan

Do you need service?

If CAL E persists even with calibration using the correct, certified weight, the balance may need to be adjusted by service.

Models showing this code: FX-GD

Source: A&D official user manual · Open manufacturer document