Code on display EC,E00 A&D FX-GD

Communication protocol error

A protocol error occurred during communication. The communication settings between the balance and the computer/printer may not match. This code is not shown on the display. It appears as a response sent over RS-232C.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Compare and match the baud rate, parity, and data format settings of the balance and the connected device.
  2. Check both ends of the communication cable and connect them firmly.

Possible causes

  • Baud rate, parity, or data format mismatch
  • Faulty or loose communication cable

Do you need service?

This is a settings mismatch. If it persists after the settings are matched and the cable is replaced, the interface may need to be checked by service.

Models showing this code: FX-GD

Source: A&D official user manual · Open manufacturer document