Code on display Start value too small Precisa XM 120

Start sample too small

According to the manual, the sample weight must be greater than 0.200 g. Drying does not start with less sample because the moisture result cannot be calculated reliably.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Place more than 0.200 g of sample on the pan.
  2. Spread the sample on the pan in a thin, even layer.
  3. Restart the drying process.

Possible causes

  • Sample weight below 0.200 g
  • Sample not placed properly on the pan

Do you need service?

The message clears when the measurement is started with enough sample. No service is needed.