Start sample too small
According to the manual, the sample weight must be greater than 0.200 g. Drying does not start with less sample because the moisture result cannot be calculated reliably.
You can usually fix this yourself.
What should you do?
Try the steps in order.
- Place more than 0.200 g of sample on the pan.
- Spread the sample on the pan in a thin, even layer.
- Restart the drying process.
Possible causes
- Sample weight below 0.200 g
- Sample not placed properly on the pan
Do you need service?
The message clears when the measurement is started with enough sample. No service is needed.
Source: Precisa official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.