Battery empty or linearization required
The battery is empty or the balance needs to be linearized again. The manual recommends first replacing the battery or connecting the device to mains power.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Replace the battery or connect the balance to mains power with the adapter.
- Turn the balance back on.
Possible causes
- Battery discharged
- Linearization corrupted
Do you need service?
If Err 11 continues while connected to mains power, a new linearization is needed. Contact technical service for this.
Models showing this code: 165 BJ
Source: Precisa official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.