Above initial range
The scale detects a load on the pan or a non-original pan at power-up and cannot take the initial zero.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Remove the load from the pan before turning the scale on.
- Check that the original OHAUS pan is installed.
- Turn the balance back on.
Possible causes
- There was a load on the weighing pan at switch-on
- Non-original or wrong pan
Do you need service?
If it continues with the original pan and an empty pan, contact service to check the load cell.
Source: OHAUS official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.