Power-on zero range exceeded
The analyzer sees too much weight on the pan support at start-up and cannot set zero. Usually a sample or sample pan was left in place.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Remove the sample and the sample pan from the pan support.
- Switch the device off and on again.
- The heater and sample chamber may be very hot. Take care when opening the cover.
Possible causes
- Sample left on the pan support at start-up
- Sample pan left on the pan support
- Load cell fault
Do you need service?
If Err 8.1 remains with an empty pan support, the load cell must be checked by a technician.
Models showing this code: MB62
Source: OHAUS official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.