Under ınitial zero range
The analyzer does not see the expected weight of the pan support. Usually the pan support is missing or not seated correctly.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Turn the device off.
- Install the pan support correctly.
- Turn the device on again.
- The heater and sample chamber may be very hot. Take care when opening the cover.
Possible causes
- Pan support not installed
- Pan support seated wrongly
- Load cell fault
Do you need service?
If Err 8.2 appears with the pan support correctly installed, the load cell must be checked by a technician.
Models showing this code: MB32
Source: OHAUS official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.