Code on display Err 8.2 OHAUS MB32

Under ınitial zero range

The analyzer does not see the expected weight of the pan support. Usually the pan support is missing or not seated correctly.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Turn the device off.
  2. Install the pan support correctly.
  3. Turn the device on again.
  • The heater and sample chamber may be very hot. Take care when opening the cover.

Possible causes

  • Pan support not installed
  • Pan support seated wrongly
  • Load cell fault

Do you need service?

If Err 8.2 appears with the pan support correctly installed, the load cell must be checked by a technician.

Models showing this code: MB32

Source: OHAUS official user manual · Open manufacturer document