Message on display CLOCK SETTING IS NOT COMPLETED YET Ishida Uni-5 / Uni-7

Clock not set

The packing and expiry dates printed on labels are calculated from the device clock. Labels with correct dates cannot be printed until the clock is set.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Press the [MODE] key and enter the ADJUST section on the display.
  2. Enter the current date and time in the TIME / DATE SET field and save.
  3. Return to the main screen and check that the message has disappeared.

Possible causes

  • The clock was never set
  • Clock reset after the backup battery ran out

Do you need service?

If the clock resets at every startup after being set, service is required to check the backup battery.

Models showing this code: Uni-5, Uni-7

Source: Ishida official user manual