Memory error
Appears as "FILE MISSING" in the user manual, which recommends clearing the RAM data and contacting the dealer. According to the service manual, the number of automatic resets is recorded in the case of Error 40.
This fault needs a technician.
What should you do?
You can safely do these while waiting for service.
- Press the [CLEAR] key. If the error recurs, contact service.
- Clearing the RAM erases stored data. Do not perform it without a backup.
Possible causes
- Corrupted memory data
- Backup battery problem
- Electrical noise
Do you need service?
Memory clearing and data restore must be done with service support.
Models showing this code: BC-4000, BC-4000L2, BC-4000L2B, BC-4000L2H
Source: Ishida official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.