Negative weight limit exceeded
The message appears when the displayed negative weight is equal to or greater than 9 divisions (9d). It is usually seen when the zero point was set with the platter loaded or when the platter has shifted.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Check that the platter is seated correctly.
- With the platter empty, press the REZERO (zeroing) key.
- If the problem persists, turn the scale off and on again.
Possible causes
- Product on the platter during zeroing
- Removal of a container that had been tared
- Platter dislodged
Do you need service?
If UF remains after zeroing and restarting with an empty platter, contact technical service for zero adjustment/calibration.
Models showing this code: DS-700, DS-700B, DS-700E
Source: DIGI / Teraoka official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.