Zero error at startup
The manual gives this code in three cases: a load on the pan at startup, the pan not installed at startup, and the reading dropping below zero after startup.
You can usually fix this yourself.
What should you do?
Try the steps in order.
- Empty the weighing pan.
- Install the pan back in place correctly.
- If it occurred after startup, install the pan and press the <C> key. Otherwise, turn the scale off and on.
Possible causes
- Product on the pan at power-on
- The weighing pan is not installed
- Pan removed after startup
Do you need service?
The error clears when the scale is restarted with the pan empty and correctly installed.
Models showing this code: BC II 100, BC II 200, BC II 800
Source: Bizerba official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.