Code on display Ht Err A&D ML50

Temperature control error

The unit could not keep the drying temperature at the set value. The manual advises leaving it off for more than half an hour and checking again.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Turn the power switch off and leave the unit off for at least half an hour.
  2. Check that the unit's ventilation is not blocked.
  3. Turn the unit on and run the self-check.
  • The heater and sample pan may be very hot, do not touch them until cool.

Possible causes

  • Heater lamp fault
  • Temperature sensor failure
  • Unit overheated

Do you need service?

If Ht Err appears after the wait, the heater and sensor must be checked by a technician.

Source: A&D official user manual · Open manufacturer document