Code on display EC,E03 A&D GX-L

Communication timeout

With the timeout setting on, the next character of the command did not arrive within about 1 second. The command is not processed because it is incomplete. This response is sent over RS-232C.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Send the command all at once, without pauses.
  2. Check the communication cable and connection.
  3. If necessary, check the timeout (t-UP) setting in the function table.

Possible causes

  • Command characters sent slowly/in pieces
  • Communication cable or connection problem

Do you need service?

This is a communication error. It is resolved by sending the command without interruption.

Models showing this code: GX-L

Source: A&D official user manual · Open manufacturer document