Code on display EC,E04 A&D FZ-GD

Command character length error

The command sent to the balance contains more characters than allowed, and the balance does not process it. This response is sent over RS-232C.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Compare the command sent with the format in the manual.
  2. Correct the extra characters and the line terminator setting, then send the command again.

Possible causes

  • Extra characters added to the command
  • Incorrect line terminator (CR LF) setting

Do you need service?

This is a command format error. No service is needed.

Models showing this code: FZ-GD

Source: A&D official user manual · Open manufacturer document