Code on display 25 PCS A&D FX-iWP

Unit weight too low

The single-piece weight calculated in counting mode is too low, and counting with this reference would cause counting errors. For a correct result, the balance asks you to bring the number of pieces on the weighing pan up to 25.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Bring the number of pieces on the weighing pan up to the 25 shown on the display.
  2. Press the PRINT key to store the reference.
  3. If you press the PRINT key without adding pieces, the balance goes into counting mode. For a correct result, add pieces and then store.

Possible causes

  • Single-piece weight too small relative to the balance resolution
  • Insufficient number of pieces used for the reference

Do you need service?

The warning clears when enough pieces are added and the reference is stored again. Service is not required.

Models showing this code: FX-iWP

Source: A&D official user manual · Open manufacturer document