Code on display rtc PF A&D FX-i

Clock backup battery error

The date and time information was lost because the internal clock's backup battery is completely discharged. Weighing performance is normal while the balance is on or in standby. Only the date/time (GLP records) is affected.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Press any key to dismiss the warning.
  2. Set the date and time again.
  • Do not open the device to replace the backup battery.

Possible causes

  • Clock backup battery depleted
  • Device left without power for a long time

Do you need service?

Weighing is not affected. If the warning repeats often, the clock backup battery must be replaced by technical service.

Models showing this code: FX-i

Source: A&D official user manual · Open manufacturer document